Remote control commands
R&S
®
LCX Series
195User Manual 1179.2260.02 ─ 02
Example: Selecting the measurement function and setting the impedance range
The example represents the remote control commands for the measurement functions.
// *************************************************
// Select the measurement function by setting
// the type of component sample.
// Select the parameters to be measured,
// e.g., parallel inductance, dissipation
FUNCtion:MEASurement:TYPE L //impedance
FUNCtion:IMPedance[:TYPE] LPD
// *************************************************
// Activate automatic range selection
FUNCtion:IMPedance:RANGe:AUTO 1
// *************************************************
// Set the impedance and DC resistance ranges manually
FUNCtion:IMPedance:RANGe:AUTO 0
FUNCtion:IMPedance:RANGe[:VALue] 10E6 // 10KOhm
// *************************************************
// Query the measurement type and ranges
FUNCtion:IMPedance:RANGe:AUTO?
// Response: 0
FUNCtion:IMPedance:RANGe[:VALue]?
// Response: 10E6
// *************************************************
// Lock the selected measurement range
FUNCtion:IMPedance:RANGe:HOLD ON
FUNCtion:IMPedance:RANGe:HOLD?
// Response: 1
// *************************************************
// Set source impedance to LowZ mode,
// i.e. set the output impedance 10 Ohm
FUNCtion:IMPedance:SOURce LOW
// *************************************************
// Query the output impedance setting
FUNCtion:IMPedance:SOURce?
// Response: 'LOW'
// *************************************************
// Set the transformer turn ratio range
FUNCtion:TRANsformer:RANGe[:TYPE] N500