EasyManua.ls Logo

R&S LCX Series - Correction:open[:Execute]; Correction:short:state; Correction:short[:Execute]; Correction:spotSpot:Frequency

R&S LCX Series
242 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
LCX Series
184User Manual 1179.2260.02 ─ 02
Parameters:
<Open Correction
State>
1 | 0 | ON | OFF
*RST: 0|OFF
Example: See Example "Setting correction prameters" on page 181.
Manual operation: See "Open/Short/Load Correction Enabled" on page 79
CORRection:OPEN[:EXECute]
Executes an open correction on all frequencies.
Example: See Example "Setting correction prameters" on page 181.
Usage: Event
Manual operation: See "Start Short/Open Correction (Full)" on page 79
CORRection:SHORt:STATe
<Short Correction State>
Activates the short correction function.
Parameters:
<Short Correction
State>
1 | 0 | ON | OFF
*RST: 0|OFF
Example: See Example "Setting correction prameters" on page 181.
Manual operation: See "Open/Short/Load Correction Enabled" on page 79
CORRection:SHORt[:EXECute]
Executes a short correction on all frequencies.
Example:
See Example "Setting correction prameters" on page 181.
Usage: Event
Manual operation: See "Start Short/Open Correction (Full)" on page 79
CORRection:SPOT<Spot>:FREQuency <Frequency>
Defines the working point for spot correction.
Assign the working point number and the frequency.
Suffix:
<Spot>
.
1..n
Parameters:
<Frequency> Default unit: Hz
Example: See Example "Setting correction prameters" on page 181.
CORRection subsystem