Remote control commands
R&S
®
LCX Series
184User Manual 1179.2260.02 ─ 02
Parameters:
<Open Correction
State>
1 | 0 | ON | OFF
*RST: 0|OFF
Example: See Example "Setting correction prameters" on page 181.
Manual operation: See "Open/Short/Load Correction Enabled" on page 79
CORRection:OPEN[:EXECute]
Executes an open correction on all frequencies.
Example: See Example "Setting correction prameters" on page 181.
Usage: Event
Manual operation: See "Start Short/Open Correction (Full)" on page 79
CORRection:SHORt:STATe
<Short Correction State>
Activates the short correction function.
Parameters:
<Short Correction
State>
1 | 0 | ON | OFF
*RST: 0|OFF
Example: See Example "Setting correction prameters" on page 181.
Manual operation: See "Open/Short/Load Correction Enabled" on page 79
CORRection:SHORt[:EXECute]
Executes a short correction on all frequencies.
Example:
See Example "Setting correction prameters" on page 181.
Usage: Event
Manual operation: See "Start Short/Open Correction (Full)" on page 79
CORRection:SPOT<Spot>:FREQuency <Frequency>
Defines the working point for spot correction.
Assign the working point number and the frequency.
Suffix:
<Spot>
.
1..n
Parameters:
<Frequency> Default unit: Hz
Example: See Example "Setting correction prameters" on page 181.
CORRection subsystem