General instrument settings
R&S
®
LCX Series
119User Manual 1179.2260.02 ─ 02
How to: see "Assigning the touch lock function to the "User Button""
on page 119
● "Open Correction", starts calibration on all frequencies for open test fixtures
contacts.
How to: see Chapter 6.2, "Configuring the test signal", on page 57
● "Short Correction", starts calibration on all frequencies for short-circuit test fix-
tures contacts.
How to: see Chapter 6.2, "Configuring the test signal", on page 57
● "Open/Short Correction", starts calibration on all frequencies for open/short-cir-
cuit test fixtures contacts.
How to: see Chapter 6.2, "Configuring the test signal", on page 57
9.3.2 Using the user button function
Depending on the assigned function, pressing the [★ (User)] key executes the action.
The following instructions show how to assign a function, given by the example of the
touchscreen lock. It is also explained how to undo the assignment.
Assigning the touch lock function to the "User Button"
To assign the function to the "User Button":
1. To access the user button settings, select [settings] > "Device" > "User Button".
2. Select "User Button Action".
3. Select "Touch Lock".
4. Confirm with "Select".
5. To activate the touchscreen lock, press the [★ (User)].
A status message informs you that the touch input is disabled.
6. To close the message, press the [Back] key.
Now you can control the instrument only with the front panel keys and navigation
controls.
Activating the locked touch input
To unlock the touchscreen:
► Press the [★ (User)].
The R&S LCX enables the touchscreen input.
9.4 Screenshot
With the screenshot function, you can capture the image of the current screen. The
R&S LCX assigns a unique, generic filename and saves the file in *.png format.
Screenshot