Remote control commands
R&S
®
LCX Series
183User Manual 1179.2260.02 ─ 02
Parameters:
<Cable Length> Range: 0 m to 1 m
*RST: 0 m
Default unit: m
Example: See Example "Setting correction prameters" on page 181.
Manual operation: See "Cable Length" on page 76
See "Test Signal Type" on page 76
CORRection:LOAD:STATe <Load Correction State>
Activates the load correction function.
Parameters:
<Load Correction
State>
1 | 0 | ON | OFF
*RST: 0|OFF
Example: See Example "Setting correction prameters" on page 181.
Manual operation: See "Open/Short/Load Correction Enabled" on page 79
CORRection:LOAD:TYPE [<Load Correction Type>]
Selects the measurement function for the load correction.
Parameters:
<Load Correction
Type>
CPD | CPQ | CPG | CPRG | CSD | CSQ | CSRS | LPD | LPQ |
LPG | LPRP | LSD | LSQ | LSRS | RX | ZTD | ZTR | GB | YTD |
YTR
CPD | CPQ | CPG | CPRG | CSD | CSQ | CSRS
Capacitive measurement type: Cp (parallel capacitance), Cs
(serial capacitance), Q (quality factor), G (conductance), D (dis-
sipation factor), R (resistance), Rp (parallel resistance), Rs
(serial resistance)
LPD | LPQ | LPG | LPRP | LSD | LSQ | LSRS
Inductive measurement type: Lp (parallel inductance), Ls (serial
inductance), Q (quality factor), G (conductance), D (dissipation
factor), R (resistance)
RX | ZTD | ZTR | GB | YTD | YTR
Resistance measurement type: R (resistance), X impedance, G
(conductance), Z (impedance), Y (admittance), TD (phase angle
degree), TR (phase angle rad)
Example:
See Example "Setting correction prameters" on page 181.
CORRection:OPEN:STATe <Open Correction State>
Activates the open correction function.
CORRection subsystem