What to do if Riele Measuring Instruments shows error at pump calibration?
- AAudrey JimenezAug 29, 2025
If the Riele Measuring Instruments displays an error during pump calibration: - Clamp pump tube. - Control volume of aspiration. - Control connectors of tubes. - Check the plug of the bubble detector for bad contact. - Check the cable of the bubble detector for bad contact. - Adjust bubble detector (OM chapter 8.2.2). - Replace inlet tube (OM chapter 9.7).