Reference Manual Rev 1.13
Page 124 004R-646-113
Name Address Type Description
00010000
H
: Underload
00008000
H
: Error (see System Error)
00004000
H
: SETUP menus active
00002000
H
: Calibration in progress
00001000
H
: Motion
00000800
H
: Centre of Zero
00000400
H
: Zero
00000200
H
: Net
For calibration, bit 13 (00002000
H
) is high
when a calibration is taking place.
Example:
Send (status):20110021;
Response (not calibrating):
81110021:00008400
Or
Response (calibrating): 81110021:0000A400
System Error 0022
H
34
d
Number Diagnostic Errors
Absolute mV/V 0023
H
35
d
Number Absolute mV/V reading where 10000 =
1.0mV/V
Unused 0024
H
36
d
Number
Gross/Net Weight 0025
H
37
d
Number These registers return weight data.
Read Final: 8 character Hexadecimal
number. Example: 00000064 for 100 kg
Read Literal: Formatted string including
decimal point units and Gross/Net indication.
Example: ā€œ 10.0 kg Nā€
Gross Weight 0026
H
38
d
Number
Net Weight 0027
H
39
d
Number
Tare Weight 0028
H
40
d
Number
Peak Hold 0029
H
41
d
Number
Manual Hold 002A
H
42
d
Number
Grand Total 002B
H
43
d
Number
Alternate Units Gross 002C
H
44
d
Number
Raw ADC counts 002D
H
45
d
Number 2,560,000 = 1.0mV/V
Alternate Units Net 002E
H
46
d
Number as above
System Fullscale 002F
H
47
d
Number Fullscale weight of the instrument.
Traceable weight
available flag
0030
H
48
d
Number 0: No traceable weights since start up
1: Traceable weight data is valid