EasyManua.ls Logo

Rinstrum R427 - Sfn: UNITS (Unit Switching Functions); Sfn: HOLD

Rinstrum R427
142 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Reference Manual Rev 1.13
004R-646-113 Page 69
12.4.7. SFn: UNITS (Unit Switching Functions)
Path Description
FUNC
└ SFn
TYPE : UNITS
KEY
MODE
UNIT
(*)
U.STR
(*)
Unit Switching enables the display and printing of
alternative units to those used for the primary
calibration of the instrument.
KEY: Select key or external input to use.
MODE: Sets the unit switching mode. Options are:
kg/lb (default): The instrument will convert
kilograms to pounds or pounds to kilograms
(depending on the primary unit).
CUSTOM: The instrument will convert
primary units to a custom unit defined by an
entered conversion factosr.
UNIT (Alternative Unit Annunciator): Set the
symbols to use for alternative units on the
instrument display. Options are:
N: Useful for Newtons of Force.
ARROW.U: Upper unit arrow
P: useful for Pints.
L: lower case ‘l’ for litres.
ARROW.L: Lower unit arrow
U.STR (Unit String): Four character alternative units
string. Used in printing alternative units.
KEY Values <OPT>
None
(Default)
, F1 .. F3,
IO1 .. IO32
MODE Values <OPT>
kg/lb
(Default)
CUSTOM
UNIT Values <OPT>
NONE
(Default)
N
ARROW U
P
L
ARROW L
U.STR Values <STR>
4 character string
12.4.8. SFn: HOLD
Path Description
FUNC
└ SFn
TYPE : HOLD
KEY
The hold key/input implements a manual hold.
KEY: Select key or external input to use.
KEY Values <OPT>
None
(Default)
, F1 .. F3,
IO1 .. IO32

Table of Contents

Related product manuals