Reference Manual Rev 1.13
Page 112 004R-646-113
Not setpoint 6 14
Not setpoint 5 13
Not setpoint 4 12
Not setpoint 3 11
Not setpoint 2 10
Not setpoint 1 9
Setpoint 8 8
Setpoint 7 7
Setpoint 6 6
Setpoint 5 5
Setpoint 4 4
Setpoint 3 3
Setpoint 2 2
Setpoint 1 1
DLY.ON: delay before setpoint becomes active.
HLD.OFF: delay before setpoint becomes inactive.
17.7. Scale Entry/Exit Setpoint Types
The following setpoint types are all based on the status of the indicator.
Scale Ready (SC.REDY): Setpoint is active when in the zero band and
stable for longer than the time set in RDY.TIM.
Scale Exit (SC.EXIT): Setpoint is active when outside of the zero band and
either a print has occurred or the scale exit (SC.EXIT) special function has
been triggered.
Ready Time (RDY.TIM): Time in seconds that the scale must be stable in the zero
band before the scale is ready.