13.4. Calibration Errors
Following are a list of the possible error messages that may be displayed to warn of
failed or incorrect calibration:
Error Description Resolution
(FAILED)
(BAND)
An attempt has been made to
calibrate with a weight or
signal which is not in the valid
range.
Check weights and retry.
(FAILED)
(ERROR)
An attempt has been made to
calibrate while the scale
signal is not valid.
Check loadcell connection and
the 4-wire/6-wire setting.
(FAILED)
(TIMEOUT)
For an unknown reason, the
calibration was unable to
complete.
Retry.
(FAILED)
(RES)
An attempt has been made to
calibrate the scale to a
resolution which is too high
for the instrument.
Check weights and retry.
(FAILED)
(TOO CLOSE)
An attempt has been made to
add a linearisation point too
close to zero, span or another
linearisation point.
Check weights and retry.
Table 5: Calibration errors