12.3.  SCALE (Loadcell options and calibration) 
12.3.1.  BUILD (Scale parameters) 
Path  Description 
SCALE 
└ BUILD 
└ TYPE
()
 
└ CABLE
()
 
└ DP
()
 
└ CAP1
()
 
└ E1
()
 
└ CAP2
(*)
 
└ E2
(*)
 
└ UNITS
()
 
└ HI.RES
() 
Scale Base configuration settings: 
TYPE: Range type.  Options are: 
  SINGLE : Single range 
  DUAL.I: Dual interval 
  DUAL.R: Dual range 
CABLE: 6-wire or 4-wire cable termination: 
  6-wire: SENSE lines are connected to the 
instrument. 
  4-wire: Internal connection between 
Excitation and SENSE lines is active. 
DP: Set the decimal point position. 
CAP1: Sets the fullscale capacity for the scale.  If 
using multiple interval/range, this sets the fullscale 
capacity of the lowest range/interval. 
E1: Sets the count-by (or resolution) of the scale.  If 
using multiple interval/range, this sets the count-by 
(or resolution) of the lowest range/interval. 
CAP2: If using multiple interval/range, this sets the 
fullscale capacity of the highest range/interval. 
E2: If using multiple interval/range, this sets the 
count-by (or resolution) of the highest range/interval. 
UNITS: Sets the weighing units. 
NB: For Options: 
  None: Units are left blank. 
  ARROW.U: Use the top arrow.  Units 
will be printed onto the instrument in the 
correct location. 
HI.RES: Sets the scale to high resolution (x10) 
mode. 
 
: This item is trade critical and will affect the 
calibration counter(s) if changed. 
TYPE Values 
()
 <OPT> 
SINGLE  
(Default)
 
DUAL.I , DUAL.R 
CABLE Values 
()
 <OPT> 
6 WIRE  
(Default)
, 4 WIRE 
DP Values 
()
 <OPT> 
000000 
(Default)
 
00000.0 
0000.00 
000.000 
00.0000 
0.00000 
CAP1 & CAP2 Values 
()
 
<NUM> 
100 ..999999 Default: 3000 
NB: Numbers above assume 
no decimal point. 
E1 & E2 Values 
()
 <OPT> 
  1  
(Default)
 
  2 
  5 
10 
  20   
  50 
100 
 
UNITS Values 
()
 <OPT> 
None 
kg  
(Default)
 
lb 
t 
g 
Oz 
N 
ARROW U 
HI.RES Values 
()
 <OPT> 
OFF  
(Default)
, ON