12.3. SCALE (Loadcell options and calibration)
12.3.1. BUILD (Scale parameters)
Path Description
SCALE
└ BUILD
TYPE
()
CABLE
()
DP
()
CAP1
()
E1
()
CAP2
(*)
E2
(*)
UNITS
()
HI.RES
()
Scale Base configuration settings:
TYPE: Range type. Options are:
SINGLE : Single range
DUAL.I: Dual interval
DUAL.R: Dual range
CABLE: 6-wire or 4-wire cable termination:
6-wire: SENSE lines are connected to the
instrument.
4-wire: Internal connection between
Excitation and SENSE lines is active.
DP: Set the decimal point position.
CAP1: Sets the fullscale capacity for the scale. If
using multiple interval/range, this sets the fullscale
capacity of the lowest range/interval.
E1: Sets the count-by (or resolution) of the scale. If
using multiple interval/range, this sets the count-by
(or resolution) of the lowest range/interval.
CAP2: If using multiple interval/range, this sets the
fullscale capacity of the highest range/interval.
E2: If using multiple interval/range, this sets the
count-by (or resolution) of the highest range/interval.
UNITS: Sets the weighing units.
NB: For Options:
None: Units are left blank.
ARROW.U: Use the top arrow. Units
will be printed onto the instrument in the
correct location.
HI.RES: Sets the scale to high resolution (x10)
mode.
: This item is trade critical and will affect the
calibration counter(s) if changed.
TYPE Values
()
<OPT>
SINGLE
(Default)
DUAL.I , DUAL.R
CABLE Values
()
<OPT>
6 WIRE
(Default)
, 4 WIRE
DP Values
()
<OPT>
000000
(Default)
00000.0
0000.00
000.000
00.0000
0.00000
CAP1 & CAP2 Values
()
<NUM>
100 ..999999 Default: 3000
NB: Numbers above assume
no decimal point.
E1 & E2 Values
()
<OPT>
1
(Default)
2
5
10
20
50
100
UNITS Values
()
<OPT>
None
kg
(Default)
lb
t
g
Oz
N
ARROW U
HI.RES Values
()
<OPT>
OFF
(Default)
, ON