Reference Manual Rev 1.13
Page 18 004R-646-113
4. Setup Menus
Throughout the setup menus different data entry methods are used. Each method is
described below.
4.1. Accessing Setup Menus
There are two methods to access the Setup area:
The Full Setup method provides access to all functions in Setup, including legal for
trade and calibration sensitive settings. Changes in Full Setup mode may result in
the calibration counter being incremented. If an attempt is made to enter Full Setup
using the incorrect passcode, the instrument will respond with the message ENTRY
DENIED. Refer to Passcodes page 53 for more information.
Full Setup
WARNING
All items in all menus will
be enabled in Full Setup.
Care should be taken to
avoid inadvertently
altering the Build or
Calibration settings.
There are 2 methods of accessing full
setup:
1. Press and hold the
<POWER> and
<F3> keys together
for two seconds, or
2. Press the setup button on the rear of
the instrument.
Safe Setup restricts access to the Trade Critical settings. Changes made in this
mode will not increment the calibration counter. In this manual, items marked with
indicate that the setting is trade critical. If an attempt is made to enter Safe Setup
using the incorrect passcode, or if an attempt is made to alter a trade critical setting
while in Safe Setup, the instrument will respond with the message ENTRY DENIED.
Refer to Passcodes page 53 for more information.
Safe Setup
Press and hold both
the <POWER> and
<ZERO> keys
together for two seconds.