Cedex HiRes Analyzer Operator’s Guide
G
G
Settings in the section “Measurement“:
Option Value Description Save last
value
AskBeforeMultiRun 0 Require to confi rm the start of a multi run
CellDiameter
HistogramDividers
(0) (0) (0) divider settings for use in the cell diameter histogram ×
CellType Std. Size Default Cell Type for measurements. ×
CellTypeFocus
Offset
-25 Default focus offset for the creation of new Cell Types if
the new Cell Type is not based on a previous Cell Type.
CellType
Sedimentation
Duration
45 Default sedimentation time (in seconds) for the creation
of new Cell Types if the new Cell Type is not based on a
previous Cell Type.
DefaultCombined
Number
1 Number of images that should be added together
for a summary of the measurements results in the
„Measurement Statistics“ histogram.
Mask image
required
0 Mask image required
Measurement
Precision
Normal
(or Medium, in
the case of a
system version
update)
Default precision for the measurement :
minimum/normal/superior (or low/medium/high in the
case of a system version update)
×
Measurement
PrescanRequired
1 Use of the Prescan in the case of the Cedex HiRes
System (0: don’t run, 1: run).
ObjectDiameter
HistogramDividers
(0) (0) (0) divider settings for use in the object diameter histogram ×
PercentageOf
ValidImages
75.0 Percentage value of images which must be valid to
ensure the measurement is valid.
×
ReactorID Default ReactorID for measurements. ×
SampleID Default SampleID for measurements. ×
SampleInitial
Dilution
1:1 Default dilution factor for measurements. (1:2, 1:20 etc.) ×
SampleVolume 300 Default sample volume for measurements
(currently not adjustable).
×
SuggestNext
CupPosition
1 Increase in cup position after every measurement,
for example when running new measurements in the
Measurement window.
(only with Auto/Multi Samplers). Possible values:
0=don’t increase, 1= automatically increase.
Workarea standard Default Workarea for measurements. ×
System Options und User Preferences