149
Appendix
G
G
Settings in the section “Measurement List Default Filter“:
Option Value Description Save last
value
CellType Measurement List Default Filter for CellType ×
DataSet Measurement List Default Filter for DataSet ×
DataSetType Effective Measurement List Default Filter for DataSet:
Effective/Original/All
×
ProcessDateFrom Measurement List Default Filter for Process Date From ×
ProcessDateTo Measurement List Default Filter for Process Date To ×
ReactorID Measurement List Default Filter for ReactorID ×
ResultsToShow
InList
12, 10, 9 Result types to be displayed in measurement list window ×
SampleDrawing
DateFrom
Measurement List Default Filter for Sample Drawing
Date From
×
SampleDrawing
DateTo
Measurement List Default Filter for Sample Drawing
Date To
×
SampleID Measurement List Default Filter for SampleID ×
ShowResults 0 Show result in measurement list window ×
User Measurement List Default Filter for UserID ×
Workarea Standard Measurement List Default Filter for Workarea ×
Settings in the section “Hardware“:
Option Value Description Save last
value
ChamberHeight 0.1 Height of the measurement chamber.
Do not change this parameter without consulting
with Roche technical services!
ErrorLevelFor
BatchStopping
1 Error level at which a Multi Run will be automatically
interrupted. (1: only fatal, 2: also critical, 3: also error, 4:
all, including warning)
Multisampler
Present
1 (not adjustable) Indication of whether an Auto/Multi
Sampler is recognized.
PixelsPermmX 1250 (not adjustable) Number of pixels per millimeter in the
x-axis of the images taken.
PixelsPermmY 1250 (not adjustable) Number of pixels per millimeter in the
y-axis of the images taken.
System Options und User Preferences