Remote Commands
R&S
®
FSW
1002User Manual 1173.9411.02 ─ 19
Example:
MMEM:SEL:ALL
//Save all items (User Settings, All Traces, All Limit Lines) from
the R&S FSW.
MMEM:LOAD:STAT 1,'C:\R_S\Instr\user\TEST01'
//Reloads all items
In the "Recall" dialog, select only "User Settings" and "All Limit
Lines".
MMEM:LOAD:STAT 1,'C:\R_S\Instr\user\TEST01'
//Reloads user settings and all limit lines.
*RST
//Reset instrument.
MMEM:LOAD:STAT 1,'C:\R_S\Instr\user\TEST01'
//Selected items are retained. Reloads user settings and all limit
lines.
Restart the instrument (using the ON/OFF key on the front
panel).
MMEM:LOAD:STAT 1,'C:\R_S\Instr\user\TEST01'
// Selected items are set to default. Reloads only the user set-
tings.
Usage:
SCPI confirmed
Manual operation: See "Recall" on page 545
See "Recall in New Channel / Recall in Current Channel"
on page 549
MMEMory:LOAD:TYPE <Mode>
This command defines whether the channel-specific settings that will be loaded with
the subsequent MMEM:LOAD:STAT command will replace the current channel or acti-
vate a new channel.
Parameters:
<Mode> NEW | REPLace
NEW
The loaded settings will be activated in a new channel.
REPLace
The loaded settings will replace the currently active channel.
*RST: NEW
Managing Settings and Results