Remote Commands
R&S
®
FSW
1003User Manual 1173.9411.02 ─ 19
Example:
INST:SEL 'SPECTRUM2'
Selects measurement channel 'SPECTRUM2'.
MMEM:STOR:TYP CHAN
Specifies that channel-specific data is to be stored.
MMEM:STOR:STAT 1, 'C:\R_S\Instr\user\Spectrum'
Stores the settings from channel 'SPECTRUM2' to the file
'C:\R_S\Instr\user\Spectrum'.
MMEM:LOAD:TYPE NEW
Specifies that channel-specific settings are to be activated in a
new channel.
MMEM:LOAD:STAT 1, 'C:\R_S\Instr\user\Spectrum'
Loads the channel-specific settings from the file
'C:\R_S\Instr\user\Spectrum' to the new channel
'SPECTRUM2*'.
MMEMory:STORe:STATe 1,<FileName>
This command saves the current instrument configuration in a *.dfl file.
Secure User Mode
In secure user mode, settings that are to be stored on the instrument are stored to vol-
atile memory, which is restricted to 256 MB. Thus, a "Memory full" error may occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details see chapter 3.1.7, "Protecting Data Using the Secure User Mode",
on page 37.
For details see "Protecting Data Using the Secure User Mode" in the "Data Manage-
ment" section of the R&S FSW User Manual.
Parameters:
1
<FileName> String containing the path and name of the target file.
The file extension is .dfl.
Example:
MMEM:STOR:STAT 1,'Save'
Saves the current device settings in the file Save.dfl.
Usage: Event
SCPI confirmed
Manual operation: See "Save File" on page 549
See "Save" on page 576
MMEMory:STORe:STATe:NEXT
This command saves the current instrument configuration in a *.dfl file.
Managing Settings and Results