Remote Commands
R&S
®
FSW
1004User Manual 1173.9411.02 ─ 19
The file name depends on the one you have set with MMEMory:STORe:STATe. This
command adds a consecutive number to the file name.
Secure User Mode
In secure user mode, settings that are to be stored on the instrument are stored to vol-
atile memory, which is restricted to 256 MB. Thus, a "Memory full" error may occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details see chapter 3.1.7, "Protecting Data Using the Secure User Mode",
on page 37.
For details see "Protecting Data Using the Secure User Mode" in the "Data Manage-
ment" section of the R&S FSW User Manual.
Example:
MMEM:STOR:STAT 1,'Save'
Saves the current device settings in the file Save.dfl.
MMEM:STOR:STAT:NEXT
Saves the current device settings in the file Save_001.dfl
MMEM:STOR:STAT:NEXT
Saves the current device settings in the file Save_002.dfl
Usage: Event
Manual operation: See "Save File" on page 549
MMEMory:STORe:TYPE <Mode>
This command defines whether the data from the entire instrument or only from the
current channel is stored with the subsequent MMEM:STOR... command.
Parameters:
<Mode> INSTrument | CHANnel
*RST: INST
Example:
INST:SEL 'SPECTRUM2'
Selects measurement channel 'SPECTRUM2'.
MMEM:STOR:TYPE CHAN
Specifies that channel-specific data is to be stored.
MMEM:STOR:PEAK 'SpectrumPeaks'
Stores the peak list from channel 'SPECTRUM2' to the file
'SpectrumPeaks'.
SYSTem:PRESet
This command presets the R&S FSW.
Example:
SYST:PRES
Usage: Event
SCPI confirmed