MMEMory:STORe:LIST <FileName>
This command exports the SEM and spurious emission list evaluation to a file.
The file format is *.dat.
Secure User Mode
In secure user mode, settings that are to be stored on the instrument are stored to vol-
atile memory, which is restricted to 256 MB. Thus, a "Memory full" error may occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details see chapter 3.1.7, "Protecting Data Using the Secure User Mode",
on page 37.
For details see "Protecting Data Using the Secure User Mode" in the "Data Manage-
ment" section of the R&S FSW User Manual.
Parameters:
<FileName> String containing the path and name of the target file.
Example:
MMEM:STOR:LIST 'test'
Stores the current list evaluation results in the test.dat file.
Manual operation: See "Saving the Result Summary (Evaluation List) to a File"
on page 225
See "Saving the Evaluation List" on page 248
MMEMory:STORe:PEAK <FileName>
This command exports the marker peak list to a file.
The file format is *.dat.
Secure User Mode
In secure user mode, settings that are to be stored on the instrument are stored to vol-
atile memory, which is restricted to 256 MB. Thus, a "Memory full" error may occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details see chapter 3.1.7, "Protecting Data Using the Secure User Mode",
on page 37.
For details see "Protecting Data Using the Secure User Mode" in the "Data Manage-
ment" section of the R&S FSW User Manual.
Parameters:
<FileName> String containing the path and name of the target file.
Example:
MMEM:STOR:PEAK 'test'
Saves the current marker peak list in the file test.dat.