Common Measurement Settings
R&S
®
FSW
324User Manual 1173.9411.02 ─ 19
The "Gate Mode" Level is not supported for R&S power sensors. The signal sent by
these sensors merely reflects the instant the level is first exceeded, rather than a time
period. However, only time periods can be used for gating in level mode. Thus, the trig-
ger impulse from the sensors is not long enough for a fully gated measurement; the
measurement cannot be completed. For details on gating see chapter 6.6.1.2, "Gated
Measurements", on page 422.
For details see "How to Configure a Power Sensor as an External (PSE) Trigger"
on page 330.
6.2.3.2 Power Sensor Settings
Power sensor settings are available in the "Power Sensor" tab of the "Input" dialog
box. Each sensor is configured on a separate tab.
State............................................................................................................................325
Continuous Value Update........................................................................................... 325
Select.......................................................................................................................... 325
Zeroing Power Sensor................................................................................................ 325
Frequency Manual...................................................................................................... 326
Frequency Coupling....................................................................................................326
Unit/Scale....................................................................................................................326
Meas Time/Average....................................................................................................326
Setting the Reference Level from the Measurement (Meas->Ref)..............................327
Reference Value......................................................................................................... 327
Use Ref Lev Offset......................................................................................................327
Average Count (Number of Readings)........................................................................327
Duty Cycle...................................................................................................................327
Using the power sensor as an external trigger............................................................327
Data Input and Output