Common Analysis and Display Functions
R&S
®
FSW
466User Manual 1173.9411.02 ─ 19
"Clear Write"
Overwrite mode: the trace is overwritten by each sweep. This is the
default setting.
The "Detector" is automatically set to "Auto Peak".
"Max Hold"
The maximum value is determined over several sweeps and dis-
played. The R&S FSW saves each trace point in the trace memory
only if the new value is greater than the previous one.
The "Detector" is automatically set to "Positive Peak".
This mode is not available for statistics measurements.
"Min Hold"
The minimum value is determined from several measurements and
displayed. The R&S FSW saves each trace point in the trace memory
only if the new value is lower than the previous one.
The "Detector" is automatically set to "Negative Peak".
This mode is not available for statistics measurements.
"Average"
The average is formed over several sweeps.
The Sweep/Average Count determines the number of averaging pro-
cedures.
The "Detector" is automatically set to "Sample".
This mode is not available for statistics measurements.
"View"
The current contents of the trace memory are frozen and displayed.
"Blank"
Removes the selected trace from the display.
Remote command:
DISPlay[:WINDow<n>]:TRACe<t>:MODE on page 910
Detector
Defines the trace detector to be used for trace analysis.
For details see chapter 7.3.1.1, "Mapping Samples to Sweep Points with the Trace
Detector", on page 452.
Note: For EMI measurements, the trace detector is used for the initial peak search
only, not for the final test. The detector for the final test is configured in the EMI marker
settings, see chapter 5.13.4.1, "EMI Marker Configuration", on page 299.
"Auto"
Selects the optimum detector for the selected trace and filter mode.
This is the default setting.
"Type"
Defines the selected detector type.
Note: If the EMI (R&S FSW-K54) measurement option is installed
and the filter type "CISPR" is selected, additional detectors are availa-
ble, even if EMI measurement is not active.
For details see chapter 5.13.3.2, "Detectors and Dwell Time",
on page 291.
Remote command:
[SENSe:][WINDow<n>:]DETector<trace>[:FUNCtion] on page 912
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion]:AUTO on page 913
Hold
If activated, traces in "Min Hold", "Max Hold" and "Average" mode are not reset after
specific parameter changes have been made.