Common Analysis and Display Functions
R&S
®
FSW
512User Manual 1173.9411.02 ─ 19
The "Level" and "Frequency" or "Time" settings define the position and value of the ref-
erence point.
Alternatively, a Peak Search can be performed to set the maximum value of the
selected trace as the reference point.
If "Automatic Peak Search" is activated, a peak search is started automatically after
each sweep and the result is used as the reference point.
Remote command:
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:Y on page 952
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:X on page 951
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:MAXimum[:PEAK]
on page 951
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise:AUTO on page 957
Switching All Phase Noise Measurements Off
Deactivates phase noise measurement for all markers.
Remote command:
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise[:STATe] on page 957
7.4.3.4 n dB Down Marker
A special marker can be defined to determine a characteristic bandwidth or time span
in a measured signal. n dB down markers are configured in the "N dB Down Config"
dialog box, using the "n dB down" function.
To display the "N dB Down Config" dialog box, do one of the following:
Press the MKR FUNC key, then select the "Select Marker Function" softkey. Then
select the "n dB down" button. Select the "N dB Down Config" softkey.
In the "Overview", select "Analysis", and switch to the vertical "Marker Function
Config" tab. Then select the "n dB down" button. Select the "N dB Down Config"
softkey.
For details see chapter 7.4.1.9, "Measuring Characteristic Bandwidths (n dB Down
Marker)", on page 490.