Remote Commands
R&S
®
FSW
957User Manual 1173.9411.02 ─ 19
Useful commands for phase noise markers described elsewhere
● CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:MAXimum[:
PEAK]
● CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:X
● CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:Y
Remote commands exclusive to phase noise markers
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise:AUTO.............................................. 957
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise:RESult?........................................... 957
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise[:STATe]........................................... 957
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise:AUTO <State>
This command turns an automatic peak search for the fixed reference marker at the
end of a sweep on and off.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:DELT:FUNC:PNO:AUTO ON
Activates an automatic peak search for the reference marker in
a phase-noise measurement.
Manual operation: See "Defining a Reference Point" on page 511
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise:RESult?
This command queries the result of a phase noise measurement.
If necessary, the command activates the measurement first.
Return values:
<PhaseNoise>
Example:
CALC:DELT2:FUNC:PNO:RES?
Outputs the result of phase-noise measurement of the delta-
marker 2.
Usage: Query only
Manual operation: See "Phase Noise Measurement State" on page 511
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise[:STATe] <State>
This command turns the phase noise measurement at the delta marker position on and
off.
The reference marker for phase noise measurements is either a normal marker or a
fixed reference. If necessary, the command turns on the reference marker.
Analyzing Measurements (Basics)