EasyManua.ls Logo

Rohde & Schwarz FSW26 - Page 958

Rohde & Schwarz FSW26
1150 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands
R&S
®
FSW
958User Manual 1173.9411.02 ─ 19
The correction values for the bandwidth and the log amplifier are taken into account in
the measurement.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:DELT:FUNC:PNO ON
Switches on the phase-noise measurement with all delta mark-
ers.
CALC:DELT:FUNC:FIX:RPO:X 128 MHZ
Sets the frequency reference to 128 MHz.
CALC:DELT:FUNC:FIX:RPO:Y 30 DBM
Sets the reference level to +30 dBm
Manual operation: See "Phase Noise Measurement State" on page 511
See "Switching All Phase Noise Measurements Off"
on page 512
11.8.3.11 Band Power Marker
The following commands control the marker for band power measurements.
Using Markers
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer:MODE....................................................958
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer:RESult?..................................................958
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer:SPAN.....................................................959
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer[:STATe]..................................................959
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer:MODE <Mode>
This command selects the way the results for a band power marker are displayed.
Parameters:
<Mode> POWer
Result is displayed as a power in dBm.
DENSity
Result is displayed as a density in dBm/Hz.
*RST: POWer
Example:
CALC:MARK4:FUNC:BPOW:MODE DENS
Configures marker 4 to show the measurement results in
dBm/Hz.
Manual operation: See "Power Mode" on page 515
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer:RESult?
This command queries the results of the band power measurement.
Analyzing Measurements (Basics)