Remote Commands
R&S
®
FSW
712User Manual 1173.9411.02 ─ 19
In addition, limits must be defined and activated individually for each channel (see
CALCulate:LIMit:ACPower:ACHannel[:RELative]:STATe, CALCulate:
LIMit:ACPower:ALTernate<ch>[:RELative]:STATe, CALCulate:LIMit:
ACPower:GAP<gap>:ABSolute:STATe and CALCulate:LIMit:ACPower:
GAP<gap>[:RELative:]STATe).
Parameters:
<State> ON | OFF
*RST: OFF
Manual operation: See "Limit Checking" on page 157
See "Limit Checking" on page 168
See "Limit Checking" on page 170
See "Limit Checking" on page 172
11.5.3.6
General ACLR Measurement Settings
The following commands control the measurement algorithm.
Useful commands for the ACLR measurement described elsewhere:
● [SENSe:]POWer:NCORrection on page 845
● [SENSe:]POWer:ACHannel:PRESet on page 699
● [SENSe:]POWer:ACHannel:PRESet:RLEVel on page 699
● [SENSe:]POWer:TRACe on page 699
● CALCulate<n>:MARKer<m>:FUNCtion:POWer:MODE on page 696
Remote commands exclusive to ACLR measurement
[SENSe:]POWer:HSPeed............................................................................................... 712
[SENSe:]POWer:HSPeed <State>
This command turns high speed ACLR and channel power measurements on and off.
If on, the R&S FSW performs a measurement on each channel in the time domain. It
returns to the frequency domain when the measurement is done.
In some telecommunications standards, high speed measurements use weighting fil-
ters with characteristic or steep-edged channel filters for band limitation.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
POW:HSP ON
Manual operation: See "Fast ACLR" on page 153