Remote Commands
R&S
®
FSW
734User Manual 1173.9411.02 ─ 19
Example:
ESP:HSP ON
Manual operation: See "Fast SEM" on page 211
[SENSe:]ESPectrum<sb>:RANGe<range>:BANDwidth[:RESolution] <RBW>
This command defines the resolution bandwidth for a SEM range.
In case of high speed measurements, the resolution bandwidth has to be identical for
all ranges.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
<range> 1...30
Selects the measurement range.
Parameters:
<RBW> Resolution bandwidth.
Refer to the data sheet for available resolution bandwidths.
*RST: 30.0 kHz
Default unit: Hz
Manual operation: See "RBW" on page 211
[SENSe:]ESPectrum<sb>:RANGe<range>:BANDwidth:VIDeo <VBW>
This command defines the video bandwidth for a SEM range.
In case of high speed measurements, the video bandwidth has to be identical for all
ranges.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
<range> 1...30
Selects the measurement range.
Parameters:
<VBW> Video bandwidth.
Refer to the data sheet for available video bandwidths.
*RST: 10.0 MHz
Default unit: Hz
Manual operation: See "VBW" on page 211
[SENSe:]ESPectrum<sb>:RANGe:COUNt? <Ranges>
This command queries the number of ranges in the sweep list.