Remote Commands
R&S
®
FSW
735User Manual 1173.9411.02 ─ 19
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
Return values:
<Ranges> Number of ranges in the sweep list.
Usage: Query only
[SENSe:]ESPectrum<sb>:RANGe<range>:DELete
This command removes a range from the sweep list.
Note that
you cannot delete the reference range
a minimum of three ranges is mandatory.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
<range> 1...30
Selects the measurement range.
Usage: Event
Manual operation: See "Delete Range" on page 215
[SENSe:]ESPectrum<sb>:RANGe<range>:FILTer:TYPE <FilterType>
This command selects the filter type for an SEM range.
In case of high speed measurements, the filter has to be identical for all ranges.
The EMI-specific filter types are available if the EMI (R&S FSW-K54) measurement
option is installed, even if EMI measurement is not active. For details see chap-
ter 5.13.3.1, "Resolution Bandwidth and Filter Types", on page 290.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
<range> 1...30
Selects the measurement range.
Configuring and Performing Measurements