Remote Commands
R&S
®
FSW
937User Manual 1173.9411.02 ─ 19
11.8.3.4 Positioning the Marker
This chapter contains remote commands necessary to position the marker on a trace.
● Positioning Normal Markers .................................................................................937
● Positioning Delta Markers..................................................................................... 939
Positioning Normal Markers
The following commands position markers on the trace.
CALCulate<n>:MARKer<m>:MAXimum:AUTO..................................................................937
CALCulate<n>:MARKer<m>:MAXimum:LEFT...................................................................937
CALCulate<n>:MARKer<m>:MAXimum:NEXT..................................................................937
CALCulate<n>:MARKer<m>:MAXimum[:PEAK]................................................................ 938
CALCulate<n>:MARKer<m>:MAXimum:RIGHt..................................................................938
CALCulate<n>:MARKer<m>:MINimum:AUTO...................................................................938
CALCulate<n>:MARKer<m>:MINimum:LEFT....................................................................938
CALCulate<n>:MARKer<m>:MINimum:NEXT................................................................... 939
CALCulate<n>:MARKer<m>:MINimum[:PEAK]................................................................. 939
CALCulate<n>:MARKer<m>:MINimum:RIGHt...................................................................939
CALCulate<n>:MARKer<m>:MAXimum:AUTO <State>
This command turns an automatic marker peak search for a trace maximum on and
off. The R&S FSW performs the peak search after each sweep.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:MARK:MAX:AUTO ON
Activates the automatic peak search function for marker 1 at the
end of each particular sweep.
Manual operation: See "Auto Max / Min Peak Search" on page 500
CALCulate<n>:MARKer<m>:MAXimum:LEFT
This command moves a marker to the next lower peak.
The search includes only measurement values to the left of the current marker posi-
tion.
In the spectrogram, the vertical marker position remains the same.
Usage:
Event
Manual operation: See "Search Mode for Next Peak" on page 499
See "Search Mode for Next Peak in X Direction" on page 502
CALCulate<n>:MARKer<m>:MAXimum:NEXT
This command moves a marker to the next lower peak.
Analyzing Measurements (Basics)