Remote Commands
R&S
®
FSW
949User Manual 1173.9411.02 ─ 19
CALCulate<n>:DELTamarker<m>:SGRam|SPECtrogram:SARea <SearchArea>
This command defines the delta marker search area.
Parameters:
<SearchArea> VISible
Performs a search within the visible frames.
Note that the command does not work if the spectrogram is not
visible for any reason (e.g. if the display update is off).
MEMory
Performs a search within all frames in the memory.
*RST: VISible
Manual operation: See "Marker Search Area" on page 504
CALCulate<n>:
DELTamarker<m>:SGRam|SPECtrogram:XY:MAXimum[:PEAK]
This command moves a marker to the highest level of the spectrogram over all fre-
quencies.
Usage:
Event
Manual operation: See "Marker Search Type" on page 503
CALCulate<n>:DELTamarker<m>:SGRam|SPECtrogram:XY:MINimum[:PEAK]
This command moves a delta marker to the minimum level of the spectrogram over all
frequencies.
Usage:
Event
Manual operation: See "Marker Search Type" on page 503
CALCulate<n>:DELTamarker<m>:SGRam|SPECtrogram:Y:MAXimum:ABOVe
This command moves a marker vertically to the next higher level for the current fre-
quency.
The search includes only frames above the current marker position. It does not change
the horizontal position of the marker.
Usage:
Event
Manual operation: See "Search Mode for Next Peak in Y Direction" on page 502
CALCulate<n>:DELTamarker<m>:SGRam|SPECtrogram:Y:MAXimum:BELow
This command moves a marker vertically to the next higher level for the current fre-
quency.
The search includes only frames below the current marker position. It does not change
the horizontal position of the marker.