EasyManua.ls Logo

Rohde & Schwarz SMW-K144 - Page 125

Rohde & Schwarz SMW-K144
285 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Signal Control and Signal Characteristics
R&S
®
SMW-K144
125User Manual 1178.8013.02 ─ 06
Routing and enabling a trigger
The provided trigger signals are not dedicated to a particular connector. Trigger signals
can be mapped to one or more USER x or T/M connectors.
Use the Local and Global Connector Settings to configure the signal mapping, the
polarity, the trigger threshold and the input impedance of the input connectors.
To route and enable a trigger signal, perform the following general steps:
Define the signal source and the effect of a trigger event.
Select the "Trigger In > Mode" and "Trigger In > Source".
Define the connector where the selected signal is provided.
Use the Local and Global Connector Settings.
Settings:
Trigger Settings Common to All Basebands............................................................... 125
Trigger Mode...............................................................................................................125
Signal Duration Unit.................................................................................................... 126
Trigger Signal Duration............................................................................................... 126
Running/Stopped........................................................................................................ 126
Arm..............................................................................................................................126
Execute Trigger...........................................................................................................126
Trigger Source.............................................................................................................127
Sync. Output to External Trigger/Sync. Output to Trigger...........................................127
External / Trigger Inhibit.............................................................................................. 128
(External) Delay Unit...................................................................................................128
(Specified) External Delay/(Specified) Trigger Delay..................................................129
Actual Trigger Delay/Actual External Delay................................................................ 129
Trigger Settings Common to All Basebands
To enable simultaneous signal generation in all basebands, the R&S SMW couples the
trigger settings in the available basebands in any instrument's configuration involving
signal routing with signal addition. For example, in MIMO configuration, routing and
summing of basebands or of streams.
The icon indicates that common trigger settings are applied.
You can access and configure the common trigger source and trigger mode settings in
any of the basebands. An arm or a restart trigger event applies to all basebands, too.
You can still apply different delay to each of the triggers individually.
Trigger Mode
Selects trigger mode, i.e. determines the effect of a trigger event on the signal genera-
tion.
For more information, refer to chapter "Basics" in the R&S SMW user manual.
"Auto"
The signal is generated continuously.
"Retrigger"
The signal is generated continuously. A trigger event (internal or external) causes a
restart.
"Armed Auto"
Trigger Settings

Table of Contents

Related product manuals