Signal Control and Signal Characteristics
R&S
®
SMW-K144
128User Manual 1178.8013.02 ─ 06
"Off"
The signal output begins after elapsing of the processing time. Signal output starts
with sample 0. The complete signal is output.
This mode is recommended for triggering of short signal sequences. Short sequen-
ces are sequences with signal duration comparable with the processing time of the
instrument.
In master-slave mode, this setting ensures that once achieved, synchronization is not
lost if the baseband signal sampling rate changes.
Remote command:
[:SOURce<hw>]:BB:NR5G:TRIGger:EXTernal:SYNChronize:OUTPut
on page 262
External / Trigger Inhibit
Applies for external trigger signal or trigger signal from the other path.
Sets the duration with that any following trigger event is suppressed. In "Retrigger"
mode, for example, a new trigger event does not cause a restart of the signal genera-
tion until the specified inhibit duration does not expire.
For more information, see chapter "Basics" in the R&S SMW user manual.
Remote command:
[:SOURce<hw>]:BB:NR5G:TRIGger[:EXTernal]:INHibit on page 263
[:SOURce<hw>]:BB:NR5G:TRIGger:OBASeband:INHibit on page 264
(External) Delay Unit
Determine whatever the trigger delay is expressed in samples or directly defined as a
time period (seconds).