IIT-2000/2010/5000/5010 Instruction Manual/English
158 05/08/2015 Version No. 001
Example
MANU:ACW:VOLT 1
Sets the ACW voltage to 1 kV.
MANU:ACW:CHISet
Set
Query
Description
Sets or returns the ACW HI SET current value in milliamps. The
test must first be in ACW mode before this command can be
used.
Syntax
Query Syntax
MANU:ACW:CHISet <NR2>
MANU:ACW:CHISet?
Parameter/
Return
parameter
<NR2>
0.001 ~ 042.0 (IIT-2000/2010)
0.001 ~ 110.0 (IIT-5010/5000)
Example
MANU:ACW:CHIS 10.0
Sets the ACW HI SET current to 10 mA.
MANU:ACW:CLOSet
Set
Query
Description
Sets or returns the ACW LO SET current value in milliamps. The
LO SET value must be less than the HI SET value. The test must
first be in ACW mode before this command can be used.
The LO SET range must use the HI SET range. If all the digits in
the LO SET range are outside the HI SET range, an error will be