Sets or returns the ACW ARC current value in mA. ARC must be
enabled before the ARC current can be set. The test must first be
in ACW mode before this command can be used.
ARC current uses the same range as the HI SET value. The ARC
current is limited to 2X the HI SET value.
MANU:ACW:ARCCurrent <NR2>
MANU:ACW:ARCCurrent?
Parameter/
Return
parameter
1.000 ~ 080.0 (IIT-2000/2010)
2.000 ~ 200.0 (IIT-5010/5000)
MANU:ACW:ARCC 0.04
Sets the ACW ARC value to 0.04 mA.
Sets or returns the DCW voltage in kV. The test must first be in
DCW mode before this command can be used.
Note: A “DC Over 50W” error will result if the DCW Voltage X HI
SET value is > 50 watts (IIT-2000/2010).
Note: A “DC Over 100W” error will result if the DCW Voltage X HI