IIT-2000/2010/5000/5010 Instruction Manual/English
05/08/2015 Version No. 001 159
produced. All digits outside the HI SET range are ignored and will
not be used.
For example:
HI SET value: 12.34
LO SET value1: 0.005 error
LO SET value2: 0.053 no error
In the example above LO SET value1 will produce an error as all
digits are outside the range of HI SET. LO SET value2 will not
produce an error, but will return 0.05, not 0.053.
MANU:ACW:CLOSet<NR2>
MANU:ACW:CLOSet?
Parameter/
Return
parameter
0.000 ~ 041.9 (IIT-2000/2010)
0.000 ~ 109.9 (IIT-5010/5000)
MANU:ACW:CLOS 20.0
Sets the ACW LO SET current to 20 mA.
Sets or returns the ACW test time in seconds. The test must first
be in ACW mode before this command can be used.
Note: A “TIME ERR” will result if the Ramp Time + Test Time is ≥