IIT-2000/2010/5000/5010 Instruction Manual/English
160 05/08/2015 Version No. 001
240 seconds when the HI SET limit is over 30mA (IIT-2000/2010)
or over 80mA (IIT-5010/5000). This applies to the ACW function
only.
In special MANU mode, the TIMER can be turned off.
MANU:ACW:TTIMe {<NR2>|OFF}
MANU:ACW:TTIMe?
0.5 ~ 999.9 seconds
TIMER OFF (special MANU mode).
0.5 ~ 999.9 seconds
TIMER is OFF (special MANU mode).
MANU:ACW:TTIM 1
Sets the ACW test time to 1 second.
Sets or returns the ACW test frequency in Hz. The test must first
be in ACW mode before this command can be used.
MANU:ACW:FREQuency {50|60}
MANU:ACW:FREQuency?
Parameter/
Return
parameter