IIT-2000/2010/5000/5010 Instruction Manual/English
166 05/08/2015 Version No. 001
MANU:DCW:REF <NR2>
MANU:DCW:REF?
Parameter/
Return
parameter
0.000 ~ 010.9 (IIT-2000/2010)
0.000 ~ 020.9 (IIT-5010/5000)
MANU:DCW:REF 0.01
Sets the DCW reference to 0.01 mA.
Sets or returns the DCW ARC current value in mA. ARC must be
enabled to set the ARC current. The test must first be in DCW
mode before this command can be used.
ARC current uses the same range as the HI SET value. The ARC
current is limited to 2X the HI SET value.
MANU:DCW:ARCCurrent <NR2>
MANU:DCW:ARCCurrent?
Parameter/
Return
parameter
1.000 ~ 20.00 (IIT-2000/2010)
2.000 ~ 040.0 (IIT-5010/5000)
MANU:DCW:ARCC 10
Sets the DCW ARC value to 10mA.