IIT-2000/2010/5000/5010 Instruction Manual/English
166 05/08/2015 Version No. 001
Syntax
Query Syntax
MANU:DCW:REF <NR2>
MANU:DCW:REF?
Parameter/
Return
parameter
<NR2>
0.000 ~ 010.9 (IIT-2000/2010)
0.000 ~ 020.9 (IIT-5010/5000)
Example
MANU:DCW:REF 0.01
Sets the DCW reference to 0.01 mA.
MANU:DCW:ARCCurrent
Set
Query
Description
Sets or returns the DCW ARC current value in mA. ARC must be
enabled to set the ARC current. The test must first be in DCW
mode before this command can be used.
ARC current uses the same range as the HI SET value. The ARC
current is limited to 2X the HI SET value.
Syntax
Query Syntax
MANU:DCW:ARCCurrent <NR2>
MANU:DCW:ARCCurrent?
Parameter/
Return
parameter
<NR2>
1.000 ~ 20.00 (IIT-2000/2010)
2.000 ~ 040.0 (IIT-5010/5000)
Example
MANU:DCW:ARCC 10
Sets the DCW ARC value to 10mA.