IIT-2000/2010/5000/5010 Instruction Manual/English
05/08/2015 Version No. 001 93
Choose/Recall an Automatic Test
Background
The tester must first be put into AUTO mode to create or run
automatic tests.
Up to 100 automatic tests can saved/recalled.
Steps
1. If the tester is in MANU mode, press and hold the
MANU/AUTO key for three seconds. This will put
the tester into Auto mode.
The tester can only switch between AUTO and
MANU mode when in the VIEW status.
MANU/AUTO
I R G B A CW D CW
E
F R E Q = 0 H z 6
0
100
k V
EIV
MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
M A N U = * * * - 0 0 1
W
MA U OT N A_0A U T O = 0 0 1 - 1 0
VIEW status
MM A UN N A_ E
m A
Note
If the chosen automatic test has not yet been setup, then the
screen will be blank except for the status and mode.