IIT-2000/2010/5000/5010 Instruction Manual/English
05/08/2015 Version No. 001 161
MANU:ACW:FREQ 50
Sets the ACW test frequency to 50Hz.
Sets or returns the ACW reference value in mA. The test must
first be in ACW mode before this command can be used.
The ACW reference value must be less than the HI SET value.
The ACW reference value must use the same range as the HI
SET value.
MANU:ACW:REF <NR2>
MANU:ACW:REF?
Parameter/
Return
parameter
0.000 ~ 041.9 (IIT-2000/2010)
0.000 ~ 109.9 (IIT-5010/5000)
MANU:ACW:REF 0.01
Sets the ACW reference to 0.01 mA.