IIT-2000/2010/5000/5010 Instruction Manual/English
05/08/2015 Version No. 001 169
Parameter/
Return
parameter
<NR1>
IIT-2010 only:
1 ~ 9999 (unit = MΩ)
IIT-5010 only:
Format A: 0.001 ~ 50.00 (unit = GΩ)
Format B: 0.001G ~ 50.00G
Format C: 1M ~ 50000M
Example
(IIT-2000/2010)
MANU:IR:RLOS 10
Sets the IR LO SET resistance to 10MΩ.
Example
(IIT-5000/5010)
MANU:IR:RLOS 0.010
Sets the IR LO SET resistance to 10MΩ.
MANU:IR:TTIMe
Set
Query
Description
Sets or returns the IR test time in seconds. The test must first be
in IR mode before this command can be used.
Syntax
Query Syntax
MANU:IR:TTIMe <NR2>
MANU:IR:TTIMe?
Parameter/
Return
parameter
<NR2>
1.0 ~ 999.9 seconds
Example
MANU:IR:TTIM 1
Sets the IR test time to 1 second.