EasyManua.ls Logo

RS PRO IIT-2000 - Manu:dcw:closet

RS PRO IIT-2000
221 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
IIT-2000/2010/5000/5010 Instruction Manual/English
164 05/08/2015 Version No. 001
Example
MANU:DCW:CHIS 5
Sets the DCW HI SET current to 5mA.
MANU:DCW:CLOSet
Set
Query
Description
Sets or returns the DCW LO SET current value in milliamps. The
LO SET value must be less than the HI SET value. The test must
first be in DCW mode before this command can be used.
The LO SET range must use the HI SET range. If all the digits in
the LO SET range are outside the HI SET range, an error will be
produced. All digits outside the HI SET range are ignored and will
not be used.
For example:
HI SET value: 12.34
LO SET value1: 0.005 error
LO SET value2: 0.053 no error
In the example above LO SET value1 will produce an error as all
digits are outside the range of HI SET. LO SET value2 will not
produce an error, but will return 0.05, not 0.053.
Syntax
Query Syntax
MANU:DCW:CLOSet<NR2>
MANU:DCW:CLOSet?
Parameter/
Return
parameter
<NR2>
0.000 ~ 010.9 (IIT-2000/2010)
0.000 ~ 020.9 (IIT-5010/5000)

Table of Contents

Related product manuals