IIT-2000/2010/5000/5010 Instruction Manual/English
168 05/08/2015 Version No. 001
Parameter/
Return
parameter
<NR1>
IIT-2010 only:
2 ~ 9999 (unit = MΩ)
IIT-5010 only:
Format A: 0.002 ~ 50.00 (unit = GΩ)
Format B: 0.002G ~ 50.00G
Format C: 2M ~ 50000M
NULL
Sets the HI SET value to ∞.
Example
(IIT-2000/2010)
MANU:IR:RHIS 10
Sets the IR HI SET resistance to 10 MΩ.
Example
(IIT-5000/5010)
MANU:IR:RHIS 0.010
Sets the IR HI SET resistance to 10 MΩ.
MANU:IR:RLOSet
Set
Query
Description
Sets or returns the IR LO SET resistance value in MΩ (IIT-2010)
or GΩ (IIT5010). The LO SET value must be less than the HI
SET value. The test must first be in IR mode before this
command can be used.
Syntax
Query Syntax
MANU:IR:RLOSet<NR1>
MANU:IR:RLOSet?