IIT-2000/2010/5000/5010 Instruction Manual/English
176 05/08/2015 Version No. 001
Syntax
Query Syntax
MANU:UTILity:FAILmode {CONT|HOLD|STOP}
MANU:UTILity:FAILmode?
Parameter/
Return
parameter
CONT
Sets/returns the fail mode as continue.
HOLD
Sets/returns the fail mode as hold.
STOP
Sets/returns the fail mode as stop.
Example
MANU:UTIL:FAIL CONT
Sets the fail mode to CONT (continue).
MANU:UTILity:MAXHold
Set
Query
Description
Sets or returns the MAX HOLD setting for the current test.
Syntax
Query Syntax
MANU:UTILity:MAXHold {ON|OFF}
MANU:UTILity:MAXHold?
Parameter/
Return
parameter
OFF
Turns MAX HOLD off.
ON
Turns MAX HOLD on.
Example
MANU:UTIL:MAXH ON
Turns MAX HOLD on.