IIT-2000/2010/5000/5010 Instruction Manual/English
05/08/2015 Version No. 001 185
Example
AUTO:STEP 100
Sets the current AUTO number to 100.
AUTO<x>:PAGE:SHOW
Query
Description
Returns the Page View of the selected automatic test in the
following format:
step1:MANU number, step2: MANU number, step3….etc.
Query Syntax
AUTO<x>:PAGE:SHOW?
Parameter/
<x>
<NR1> 1~100
Example
AUTO1:PAGE:SHOW?
>01:011 ,02:004 ,03:003 ,04:014 ,
>05:015 ,06:020* ,07:012 ,08:018 ,
>09: ,10: ,11: ,12: ,
>13: ,14: ,15: ,16: ,
Shows the Page View for AUTO number 1.
AUTO:PAGE:MOVE
Set
Query
Description
Moves the source step to the desired destination.
Syntax
AUTO:PAGE:MOVE <Value1>,<Value2>
Parameter/
<Value1
>
<NR1> 1~16 (source step)

Table of Contents