If the (GB current x HI SET resistance) > 5.4V, then an error will
be generated (“GBV > 5.4V”).
Sets or returns the GB LO SET resistance value in mΩ. The LO
SET value must be less than the HI SET value. The test must
first be in GB mode before this command can be used.
MANU:GB:RLOSet<NR2>
MANU:IR:RLOSet?
Parameter/
Return
parameter
MANU:GB:RLOS 50
Sets the GB LO SET resistance to 50mΩ.
Sets or returns the GB test time in seconds. The test must first be
in GB mode before this command can be used.
MANU:GB:TTIMe <NR2>
MANU:GB:TTIMe?