How to Use the Bar Code Reader, and It's Function (Options)
2-35
How to Enter Sakura Solution Bar Code Data
1. Prepare the solution to be set to the instrument.
2. Touch the Read Barcode button on the Stain Process Menu.
3. Move to the Read Barcode screen, and confirm the station for which the special solution is set.
Only the stations for which bar code can be entered are displayed.
The bar code reading procedure and solution state vary depending on the station/text color, so use the
table below as reference.
Station color
Yellow
Sakura staining
solution station
Green
Kit staining solution
station
Solution name color
Black
Unexpired solution
Purple
Solution expiring that day
Red
Solution whose bar code
is not yet entered, or
expired solution
4. Select the applicable station and read the solution bar
code.
5. Once the bar code of the special solution has been read,
the information of the special solution is displayed on the
screen.
6. If the information displayed on the screen is correct, set
the special solution to the specified station.
The special solution pouch is easy to deform, so use
one hand to orient its solution outlet toward the
station. Also, squeezing the pouch causes the
solution to gush out, so gently hold the pouch with
the other hand.
7. Once the solution has been set, touch the Enter button.
8. Touch the Save button and then touch the Yes button in
the figure below, and the entered bar code information
will be saved.

Table of Contents

Other manuals for Sakura Tissue-Tek Prisma Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sakura Tissue-Tek Prisma Plus and is the answer not in the manual?

Sakura Tissue-Tek Prisma Plus Specifications

General IconGeneral
BrandSakura
ModelTissue-Tek Prisma Plus
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals