Daily Inspection and Cleaning
3-2
Replacing Activated Carbon Filters
When to replace the activated carbon filters can be easily determined by setting a time limit using the
fume filter management function (refer to P. 1-41 for details). Always replace the two activated carbon
filters in each filter pair together. If the above is not followed, the overall performance of the activated
carbon filters may significantly decrease. To replace the activated carbon filters, find a black knob located
in the filter storage area. Tilt down the knob while pulling it toward you to draw out the filter storage door.
After the filters have been replaced, store the filter storage door into the original position. Push in the door
until a click sound is heard. If the filter storage door is not set properly, the storage door may contact the
arm and cause problems.
Cleaning the Arm Holder and Basket Adapter
The arm holder groove and the metal part of the basket adapter become dirty due to repeated staining
operations. Use a cloth, etc., moistened with alcohol to clean the arm holder and basket adapter. Clean
them at least once a month.
Basket adapter
metal part
Arm holder groove
Activated carbon filter
* One filter is set on the right and
left, respectively.
Activated carbon filter
Open/close knob