System Setup
1-42
Adjusting the Water Flow
You can change the run-up time for water supply to the wash stations by selecting Water Flow from the
Utility Menu.
Water supply run-up time --- The time at which water supply to the wash stations is started during stain
process, prior to the transfer of tissues into the wash stations.
Change the setting whenever necessary before the start of staining, after changing the solution
configuration or staining program, and so on. The water supply run-up time must be changed after the
wash solution configuration has been changed. The water flow can be adjusted with the cover open.
1. Confirm that water has been drained from all wash stations.
2. Open the water supply handle.
3. Touch the Start key and wait for water to overflow from the wash stations.
4. Once water starts to overflow from the wash stations, touch the Stop key. Adjust the water supply
handle so that water will overflow within 20 to 30 seconds.
5. The water supply run-up time you have just set is displayed and the save screen appears. Touch Save
if you want to use this run-up time. Touching Exit returns to the Water Flow screen.
Water supply handle

Table of Contents

Other manuals for Sakura Tissue-Tek Prisma Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sakura Tissue-Tek Prisma Plus and is the answer not in the manual?

Sakura Tissue-Tek Prisma Plus Specifications

General IconGeneral
BrandSakura
ModelTissue-Tek Prisma Plus
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals