Operating Procedure
2-12
Removing a Basket during Stain Process
The following operation is performed when removing a basket from the end station or PE station during
stain process or after completion of stain process. It cannot be performed if the end station or PE station
contains no basket.
1. Confirm that the arm is not on a solution station in the frontmost row, and then open the door and
remove the basket.
2. Once the basket has been removed and the door closed, the basket removal confirmation window
appears. Touch the Yes key in this window. If you have not yet removed the basket, touch the No key.
End of Stain Process
Stained baskets are stored in the specified end station or PE station. (If the instrument is linked to the
Coverslipper, stained baskets are stored in the link station. If the link station is full, stained baskets are
temporarily stored in the end station and wait there until they can be stored in the link station.)
Removing Stained Baskets
1. A window notifying the end of stain process appears and automatically disappears in around 15
seconds.
2. Confirm that the arm is not on a solution station in the frontmost row, and then open the door and
remove the basket.
3. Once the basket has been removed and the door closed, the basket removal confirmation window
appears. Touch the Yes key in this window. If you have not yet removed the basket, touch the No key.
After the door is opened, remove all stained baskets from the end station and PE station.
End of Operation
This section explains the procedure to complete the stain process and end the operation.
1. After all operations have completed, turn off the power switch located at the bottom right of the
instrument, and then stop the water supply.
2. Close the lids on all solution reservoirs, and then close the cover.
This completes the series of operations.

Table of Contents

Other manuals for Sakura Tissue-Tek Prisma Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sakura Tissue-Tek Prisma Plus and is the answer not in the manual?

Sakura Tissue-Tek Prisma Plus Specifications

General IconGeneral
BrandSakura
ModelTissue-Tek Prisma Plus
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals