1-1
Installation Method
Installing the Instrument
This section provides information on determining a location for, and installing the instrument.
Installation should be performed by qualified service staff working exclusively with the instrument.
The instrument must be installed correctly to ensure proper operation and service.
Read this Operating Manual carefully before operating the instrument. Strictly observe all instructions
provided in this document.
This is a precision instrument and must be handled with care. If the instrument is handled in a rough
manner or dropped, its internal components may malfunction or get damaged. Always exercise due
care when handling the instrument.
Flow of Installation
Details: P. 1-2
Checking the installation conditions
Checking the work area
Transporting the instrument
Details: P. 1-4
Unpacking and installation
Installation check
Details: P. 1-16
Turning on the power
Required Tools
Name
Size
Q'ty
Transport handle
-
4
Phillips screwdriver
ISO No. 2
1
Adjustable wrench
300 mm or more
1
Wrench or box wrench
24 mm across flats
2
Convex ruler (steel tape
measure)
Capable of measuring objects of 2.5 m or larger
1
Level
-
1
Cutter
-
1
Pipe wrench
200 mm or more
1
Installation Environment
As is the case with all electronic devices, the instrument should not be installed in places subject to
excessive temperature or humidity.
Temperature and humidity should be held at constant levels to achieve the highest degree of operating
stability. The ambient operating temperature of the instrument is in a range of +10 to +40C. Its ambient
operating humidity is in a range of 30 to 85% in relative humidity (non-condensing).
Also, locate the instrument, with at least 10 cm away from walls, in a well-ventilated place not exposed to
corrosive vapor or extremely high/low temperatures or where the instrument will not come in direct contact
with air flow. Install the instrument away from direct sunlight and do not install it proximity to open
windows, ovens, open flames, hot plates, radiators, and dry ice baths. The instrument should be installed
away from any equipment carrying high voltage or current, such as a large refrigerator or oven. Since the
instrument is very heavy, the experiment bench should have sufficient bearing strength, and the
instrument should be laid flat.
The instrument should be installed near a power supply meeting the power facility requirements (voltage,
current) specified in this Operating Manual. Be sure to ground the power outlet and wire the power
properly using a dedicated cable.
If the instrument is used in a non-ventilated place, serious health risks may occur to the human body.

Table of Contents

Other manuals for Sakura Tissue-Tek Prisma Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sakura Tissue-Tek Prisma Plus and is the answer not in the manual?

Sakura Tissue-Tek Prisma Plus Specifications

General IconGeneral
BrandSakura
ModelTissue-Tek Prisma Plus
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals