Safety Precautions
(l)
CAUTION
Do not clean the solution reservoirs and baskets using hot water or heat the solution
reservoirs containing solutions in a microwave oven.
Heating the solution reservoirs and baskets using hot water or in a microwave oven may cause the
reservoirs/baskets to deform. Severely deformed baskets and solution reservoirs may rub against
each other on side faces or at corners, thereby inhibiting the operation.
Do not use any ultraviolet sterilizing lamp around the instrument.
Ultraviolet light may cause the instrument characteristics to change or deteriorate.
Fill each solution reservoir with a specified amount of solution.
Failure to do so may cause tissues to dry up or result in staining problems.
Replace the activated carbon filters in the fume control unit periodically.
Replace the activated carbon filters regularly in order to maintain a better environment around the
instrument. Expired activated carbon filters cannot maintain exhaust gas at an appropriate level of
concentration. When replacing the activated carbon filters, put the lids on the solution stations or
remove the solution stations by considering the possible scattering of activated carbon particles, to
prevent solution contamination.
* If the instrument is connected to an external exhaust device, the activated carbon filters need not be
replaced.
Use dedicated activated carbon filters for the fume control unit.
Use of inappropriate activated carbon filters may cause explosion of powder dust generated from
activated carbon powder.
Inspect the instrument every 6 months.
Maintain/inspect the instrument periodically so that it operates correctly at all times. For details on
periodic inspection, contact the Sakura Finetek sales representative.
If the instrument must be moved after installation, contact the Sakura Finetek Technical
Support representative or local representative.
Unexpected accident may occur by relocating the instrument, so contact the Sakura Finetek Technical
Support representative or local representative.
Use designated consumable parts.
When replacing consumable parts, use those specified in this manual. Use of any other consumable
part may result in malfunction or failure.
Close the lids of solution reservoirs whenever the instrument is not operating.
If the solution reservoir lids remain open for a prolonged period of time, organic gas may be released
to the room.
Provide sufficient ventilation around the instrument.
Some of the solutions used in the instrument are toxic/harmful to the human body. Accordingly,
provide sufficient ventilation around the instrument.
Install the instrument away from other electrical equipment.
Electromagnetic waves and vibration, heat and solution may affect and be affected by other systems
located nearby.

Table of Contents

Other manuals for Sakura Tissue-Tek Prisma Plus

Questions and Answers: