EasyManua.ls Logo

Sartorius PR 5220 Series - Page 109

Sartorius PR 5220 Series
172 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PR 5220 Instrument Manual Commissioning
Sartorius EN-109
2. Determining a condition
The selected action of each digital input can be combined with a condition that must be met for signal change
from 0 to 1 (on) or for signal change from 1 to 0 (off). Select the condition from the list; see page 104. No
condition is defined when selecting [no condition]; the action is executed directly.
Example: Taring via the digital input only if the gross weight exceeds the limit value.
Press
and select [Digital I/O parameter] to open the configuration menu.
Setup/Digital I/O parameter
Output 1
marker bit 1
X64=1
Output 2
marker bit 2
X65=1
Output 3
marker bit 3
X66=1
Input 1 on
set tare
X113=1
Condition
tare active
X58=0
Input 1 off
-no action-
Input 2 on
-no action-
Input 2 off
-no action-
Input 3 on
-no action-
Input 3 off
-no action-
BCD out
Gross
In this example: If input 1 changes from 0 to 1 [input 1 on], a taring signal is triggered only if the condition under
[Condition] is met (limit 1 out = active).

Table of Contents

Related product manuals