PR 5220 Instrument Manual Commissioning
Sartorius EN-95
4.6.12 Subsequent Dead Load Correction
If the vessel weight changes by an amount that is higher than the zero-setting range, e.g. due to dead load
reduction, dead load increase or mechanical changes, the functions for automatic zero tracking and manual
zero setting no longer work. To view the range which is currently being utilized for zero tracking or zero
setting, press
. This also activates 10-fold increased resolution of the weight value. Press
again to return
to the previous state:
Current zero set: 0.00001 t
If the full zero-setting range is already being utilized, you can still correct the dead load (CAL switch must be
open) without affecting other calibration data/parameters.
Weighingpoint/WP A/Calibration
Number of vessel feet
4
Select [Dead load] and see page 90.
Check the corner load, see Chapter 4.6.9.2.
Carry out a corner correction if necessary, see
Chapter 4.6.9.3.
Local gravity
9.81379 m/s
2
Max
10000 d
200.000 t
Scale interval
1 d
0.020 t
Dead load
7.512 t
CAL weight
10 t
Corner correction
ok

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius PR 5220 Series and is the answer not in the manual?

Sartorius PR 5220 Series Specifications

General IconGeneral
BrandSartorius
ModelPR 5220 Series
CategoryTransmitter
LanguageEnglish

Related product manuals